MTS Series 793 User Manual

Page 387

Advertising
background image

Description

Item

Displays the accumulated time in which the test program is running.
The Rewind control resets this value, the Stop and Hold controls do

Running Time

not. Pressing Run after pressing Stop or Hold causes this value to
resume incrementing.

Similar to Rollover Time except it increments only while the station
test state is not in the stopped state.

Rollover Running Time

Displays integer signals such as, interlocks, HSM states, hold and
run/stop states, and digital inputs and outputs. Most Integer signals

Integer Signals

are available when the scope traces’s Channel setting is set to Others.
Some integer signals appear in channel lists.

Note:

Unitless integer signals, such as HSM ON and Hold are not
available for meters.

Displays the signal that goes into the valve driver in volts.

Output

Displays various types of sensor feedback signals in engineering units.
Force and Displacement are typically sensor signals.

Sensor signals

Displays the number of segments played out so far in the selected
channel. When monitoring signals with a meter, this signal is only
available for Timed meters.

Count

Displays the command signal in engineering units. This signal can be
internally generated by the Controller or externally generated by an
analog input.

Command

Displays the command signal after going through a compensator (that
is, an Adaptive Compensator) in engineering units. This does not apply

Comp. Command

to null pacing. If null pacing is used, this signal is the same as the
command signal.

Displays the control mode error signal in engineering units. This signal
shows the difference between the Channel Name Command and the
Channel Name Control Mode Name signals.

Error

Displays the sensor feedback of the active control mode.

Active Fdbk

Displays the absolute value of the Error signal in engineering units.
Absolute means negative becomes positive and positive remains
positive. This signal is always positive.

Abs. Error

Displays command frequency.

Command Frequency

MTS Series 793 Control Software 387

Scopes and Meters

Advertising