Warning:

Changes in Function Generator values made with hydraulic pressure on can result in
unexpected actuator movement.

A moving actuator can injure anyone in its path.

Always clear the actuator area before changing values.

2. In the Function Generator panel:

a) Select the Channel to be programmed.

b) Select a Control Mode of displacement.

c) Select a Command Type of Cyclic.

d) Set a Target Setpoint of midstroke.

Typically, the actuator should cycle around midpoint.

e) Set an Amplitude (±) that will move the actuator through about 70% of its full displacement.

f) Set a Frequency of 0.25–0.5 Hz.

Displayed ranges for Target Setpoint, Amplitude, and Frequency are saved and are restored when
returning to a previously programmed channel.

g) Select a Wave Shape of Sine.

h) As needed, select a Compensator.

Warning:

Clicking the Run button will put actuators in motion.

A moving actuator can injure anyone in its path.

Always clear the test area before pressing the run button.

3. In the Station Manager window’s Station Controls panel:

a) Click Reset to clear interlocks.

b) Apply station power.

c)

Click

to start the program.