MTS Series 793 User Manual
Page 76
Advertising
Description
Callout
Command
2
PIDF
3
Valve Driver
4
Conditioner
5
Drive
6
Feedback
7
Load Frame/Actuator
8
Program with feedback channels
Program with feedback channels are used to program external controllers while monitoring feedback for
command compensation or data acquisition. In this channel configuration, the external controller maintains
closed-loop control.
Description
Callout
Controller
1
Command
2
Feedback
3
External Controller
4
PID
5
Drive
6
Program only channels
Program only channels are used to send programming to an external controller. In this channel configuration,
the external controller maintains closed-loop control.
76 MTS Series 793 Control Software
Station Builder