Current sensors – Rockwell Automation 1412-PP21xx PowerPad Portable Powermonitor User Manual

Page 19

Advertising
background image

Rockwell Automation Publication 1412-UM001D-EN-P - September 2012

19

Operation

Chapter 2

Current Sensors

Follow these steps the select the current sensor type.

1. Highlight CURRENT SENSORS with the Up/Down Cursor buttons,

then press Enter.
This screen appears.

2. Choose the sensor type with the Up/Down Cursor buttons.

The MN Clamp also requires a range selection to match your probe. The

range choices are 200, 100, and 5 A. The 200 A range is used for the

MN93 Probe and the 100 or 5 A range is used for the MN193 Probe.
a. To select the proper range, first ensure that the MN Probe choice is

highlighted by using the Up/Down Cursor buttons.

b. Press the Left Cursor button to highlight the range, then press either of

the Up/Down Cursor buttons to select the desired measurement range

of 200, 100, or 5 A.
If the 5 A range is selected, a choice for programming the ratio is

offered.

c. To change the ratio, press the Left Cursor button to highlight the

secondary value.

d. Press either of the Up/Down Cursor buttons to toggle this value to

either 1 or 5.

e. After making this selection, press the Left Cursor button to adjust the

primary value.

Table 1 - Synchronization of the Display in Waveform Mode

Display Selection

(Vertical Right Menu)

Reference Channel

for Synchronization

3U

U1

3V

V1

4A / 3A

A1

L1

V1

L2

V2

L3

V3

MN clamp

SR clamp

AmpFlex

MR clamp

Adapter

200A

1000 / 5A

07/25/02 10:26

100%

1

2

3

Current transducer ratio.

Left/Right Cursor buttons moves the cursor to

select the digit to edit.

Up/Down Cursor buttons increases or

decreases the value at the highlighted

position.

Secondary current value.

Nominal value of primary current from

5…2999 A.