Apparent Power (VA)

0VA … 9999 kVA

4 digits

±1 % ± 1 ct

Power Factor (PF, DPF)

-1.000 … 1.000

Cos

 0.5

Cos



0.2 … < 0.5

0.001

±1.5% ± 1 ct

±1.5% ± 1 ct

Tangent (Tan) for VA  50VA

-32.76 … 32.76

0.001 Tan

< 10

0.01 Tan

< 10

±1 ° on

±1° on

Active Energy (Wh)

Clamp-on and AMPFLEX
Clamp-on
AMPFLEX

0 Wh … 9999 MWh
Cos

 0.8

Cos

0.2 … < 0.8

Cos

0.5 … < 0.8

4 digits

(10000 cts)

±1 % ± 1 ct
±1.5% ± 1 ct
±1.5% ± 1 ct

Reactive Energy (VARh)

Clamp-on

0 VARh … 9999 MVARh
Sin

 0.5

Sin



0.2 … < 0.5

4 digits

±1.5% ± 1 ct
±2.5% ± 1 ct

Reactive Energy (VARh)

AMPFLEX

0 VARh … 9999 MVARh
Sin

 0.5

Sin

0.2 … < 0.5

4 digits

±1.5% ± 1 ct
±2.5% ± 1 ct

Apparent Energy (Vah)

0 VAh … 9999 MVAh

4 digits

±1 % ± 1 ct

Unbalance (Vunb, Aunb) three phase

supply

0% … 100%

0.1 %

±1 % ± 1 ct

Phase angle



(V/I; I/I; V/V)

-179° … +180°

± 2°

Harmonics ratios

F = 40…69 Hz

(Vrms > 50 V)

(rms > I nom/100)

0%… 999%

1…50th harmonic

0.1 %

± 1 % + 5 cts

Harmonics angles

F = 40…69 Hz

(Vrms > 50 V)

(rms > I max/100)

-179° … +180°

1 … 25

26 … 50

± 3°

± 10°

Total harmonics distortion

(Vthd, Athd, Uthd)

0% … 999%

50th harmonic

0.1 %

±1 % + 5 cts

Current K factor (Akf)

1 … 99.99

0.01

±5% ± 1ct

Voltage flicker (Vflk)

0.00 … 9.99

0.01

not specified

Table 4 - Accuracy Specifications 1412-PP2127

Function

Range

Display Resolution

Accuracy

IMPORTANT

Vdem, Adem and Udem = Values measured on half cycle (allow to get a
measurement with positive and negative values).
Vpm, Upm and Apm = Peak values min.