1000 amp ac and – Rockwell Automation 1412-PP21xx PowerPad Portable Powermonitor User Manual

Page 83

Advertising
background image

Rockwell Automation Publication 1412-UM001D-EN-P - September 2012

83

Specifications

Appendix B

1000 Amp AC and

1400 Amp DC Current Probe

(Base Unit Accuracy Included)

Attribute

1412-PP2137-05

Nominal Range

1000 A AC 1400 A DC max

Measurement Range

10…1000 A AC max, 10…1300 A peak AC + DC

(1)

(1) Currents < 1 A AC/DC will be displayed as zero on the PowerPad unit with this probe.

Probe Output Signal

1 mV/A

Maximum Cable Diameter

One 42 mm (1.6 in.) or two 25.4 mm

(0.98 in.) or two bus bars 50 x 5 mm

(1.96 x 0.19 in.)

Safety

EN 61010-2, 600 V, Category III and/or 300V Cat IV,

Pollution Degree 2

Attribute

1412-PP2137-05

Ambient temperature

18…28 °C (64…82 °F)

Humidity

20…75% of RH

Battery voltage

9V ±0.1V

Position of conductor in the sensor

Centered

Magnetic field

DC magnetic field

External magnetic field, ac

None

External electric field

None

Frequency

 65 Hz

Type of signal measured

Sinusoidal

Attribute

1412-PP2137-05

Primary current

10…100 A

100…800 A

800…1000 A AC

800…1300 A Peak

Accuracy

 1.5% + 1 ct

 3% + 1ct

 5% + 1 ct

Primary current

10…100 A

100…1000 A

-

Phase angle

 2.0°

 15°

-

Table 11 - Factors Affecting Accuracy (% of the Output Signal)

Conditions

Range

Error

Temperature

18…28 °C (64…82 °F)

Zero:  2 A/°C

Scale:  300 ppm/°C or 0.3%/10 °C

Relative humidity

10 … 90%

0.5% of Reading

Battery voltage

(1)

(2)

6.5 … 10 V

Position of a 20 mm Ø 20 conductor

DC at 440 Hz

DC at 1 Hz

DC at 2 Hz

DC at 5 Hz

< 0.5% of Reading

< 1 % of Reading

< 3% of Reading

< 10% of Reading

Live adjacent conductor

50 and 60 Hz

< 10 mA/AAC (1 in from clamp)

External field

400 A/m

< 1.3 A