Rockwell Automation 1412-PP21xx PowerPad Portable Powermonitor User Manual

Page 38

Advertising
background image

38

Rockwell Automation Publication 1412-UM001D-EN-P - September 2012

Chapter 3

Display Modes

This table lists the capture threshold levels, based on the probe in use (for the

current channels) and voltage at the different percent selections.

Transients are detected by comparing all 256 samples on the current cycle with

their counterparts from the previous cycle for each active input channel. Should

any one sample deviate from its counterpart by the selected percentage value in

the set up, this will be considered a transient and the data will be captured.

When capture occurs, four cycles are recorded for each input. These include the

trigger cycle, the previous cycle to the trigger and the two cycles that follow the

triggered cycle. All active inputs will be captured.

Thresholds
100%

50%

20%

10%

5%

2%

1%

MN93

200 A

100 A

40 A

20 A

10 A

4 A

2 A

MN193 (100A)

100 A

50 A

20 A

10 A

5 A

2 A

1 A

MN193 (5A)

[(primary x 5) ÷ (secondary)] x (percent x 100)

SR193

1000 A

500 A

200 A

100 A

50 A

20 A

10 A

AMPFLEX 193

2900 A

1400 A

580 A

290 A

140 A

58 A

29 A

MR193

1000 A

500 A

200 A

100 A

50 A

20 A

10 A

2999A Ratio

Adapter

3000 A

1500 A

600 A

300 A

150 A

60 A

30 A

1A Ratio Adapter

1 A

0.5 A

0.2 A

0.1 A

0.05 A

0.02 A

0.01 A

Voltage

480V

240V

96V

48V

24V

9.6V

4.8V