Metrohm 870 KF Titrino plus Manual User Manual

Page 33

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation

870 KF Titrino plus

■■■■■■■■

23

ID2

Sample identification.

Entry

max. 10 characters

Default value

empty

Sample size

Sample size.

Input range

–999999999 - 9999999999

Default value

1.0

Unit

Unit of the sample size.

Selection

g | mg | µg | mL | µL | pieces | User-defined

Default value

g

User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined.

5.4.2

Requesting sample data at the start of the determination

The sample data can be requested immediately after the start of the deter-
mination in order to ensure that the sample data entry is not forgotten.
This automatic request is indispensable when you reweigh your samples.

The corresponding parameters must be activated under Start conditions
for this purpose. If the parameter Hold at request is activated, then the
run will be paused and must be continued with [START] after the input
of the sample data. If Hold at request is deactivated, then the titration
will be started in the background. This dialog will be displayed until the
entering of the sample data is confirmed with [START], even if the the
titration is already completed. This ensures that the sample data is avail-
able for calculations.

Advertising