10 Appendix

870 KF Titrino plus

■■■■■■■■

79

Command

Function

Comment

$D

Scan instrument sta-
tus

Acknowledgements: Ready;0, Busy;0,
Hold;0 or Cond;0 (0 = no message)

If a message on the instrument requires
the interaction of the user, the acknowl-
edgement of the status scan displays the
corresponding message number. Exam-
ple:

Busy;010-119

= "Check buret unit"

The message can be confirmed with [OK]
or [Cancel], see below.

$A

Confirm message

Confirm the message on the instrument
with [OK]

A mandatory status scan providing the
message number must take place immedi-
ately before confirming the message, see
above.

$A(OK), $A(CANCEL)

Confirm message

Confirm the message with [OK] or [Can-
cel]

$A(YES), $A(NO)

Confirm message

Confirm the message with [Yes] or [No]

$L(method name)

Load method

The method name has to be known and
unique.

$Q(variable)

Request variable value

Examples for variables: EP1, R1, C00.

List of variables, see below.

Variable

Comment

C00

Sample size

EP1

Volume of the endpoint EP1

CI#

Sample identification (# = 1…2)

R1

Result

CV01

Common variable (blank value at KFT Ipol)

CV02

Common variable (blank value at KFT Upol)

SMN1

Mean value of result R#

TITER

Titer of selected solution