6.1 Basic settings

■■■■■■■■■■■■■■■■■■■■■■

40

■■■■■■■■

870 KF Titrino plus

NOTE

In order to ensure that a second language can be selected, it must first
be installed. This installation must be carried out by specialist personnel.
In chapter Language files, page 49, you will find details regarding the
installation of a second language.

Dialog type

The user dialog can be limited for routine operations. One can operate
normally with methods in the limited dialog. However, no settings can be
made or methods deleted.

The resetting of the dialog will take effect as soon as you exit the main
menu.

The limitation of the dialog results in the following:

The menu items System and Parameters are not shown in the main
menu.

Methods can only be loaded, but not deleted, exported or created.

NOTE

If the limited dialog for routine operation is activated, then the expert
dialog cannot be activated during ongoing operation. To change the
dialog type, the 870 KF Titrino plus must be switched off and then back
on again. The expert dialog can be forced as soon as the instrument is
started up again. Then it is possible to enter whatever settings one
wishes, e.g. the changing of the dialog type. If the instrument is
switched off again without changing the dialog type, then the routine
dialog will remain activated.

Forcing the expert dialog:

Switch on the instrument.

Wait for the display of the instrument logo with the lettering easy,
safe, precise
.

Press the [STOP] key once again and hold it down while also briefly
pressing the [BACK] key.

Release both keys once again.

Selection

Expert | Routine

Default value

Expert

Expert
Complete dialog.