5 Operation

870 KF Titrino plus

■■■■■■■■

25

3 Adding sample

NOTE

Calculate the amount of the sample in such a way that a titrant
consumption of 10…90% of the cylinder volume will result.

If Conditioning OK is displayed, press [START].
Conditioning is stopped. The request for adding the sample will
be displayed for 8 s.
The sample must be added during this time.

Add the sample.

Afterward, the request for the sample size appears:

4 Entering the sample size

Press [OK].
The editing dialog opens.

Enter the sample size and apply with Accept or [BACK].

5 Starting the titration

Press [START].

The titration starts and the curve is displayed: