5 Operation

870 KF Titrino plus

■■■■■■■■

33

PC/LIMS

Machine-readable report with all of the data for
a determination. This report can be saved as a
TXT file to a connected USB flash drive or sent to
a terminal program or a LIMS via an RS-232
interface. The definition is made in the system
settings (see "PC/LIMS report", page 46).

Report as in
method

The reports that are defined in the method will
be printed out.

5.10

Manual control

Menu

Manual control

The following functions are available in the manual control:

Dosing

Stirring

The available subfunctions are listed for each function in the function bar.

NOTE

The Stirrer function can only be selected if a stirrer is connected.

5.10.1

Dosing

The following dosing functions are available in the manual control:

Prepare the buret
unit (PREP)

Rinse the cylinder and tubings of the buret unit
(see Chapter 10.1.2, page 70).

Continuous dosing
(DOS)

Dose while the [START] key is pressed.

Dose a fixed vol-
ume (ADD)

Dose a specified volume.

Preparing the buret unit (PREP)
The PREP function is used to rinse the cylinder and tubings of the buret
unit and fill them air bubble-free. You should carry out this function
before the first determination or once per day.