10 Appendix

870 KF Titrino plus

■■■■■■■■

77

Outputs

Open Collector

tp

t

p

> 200 ms

active = low, inactive = high

I

C

= 20 mA, V

CEO

= 40 V

+5 V: maximum load = 20 mA

10.6.2

Status diagram of the remote interface

[START]

Output 0

Output 1

Output 3

Output 2

Output 4

ready/EOD

condOK

Error

Start

Stop

Input 0

Input 1

EOD

[START]

[STOP]

Titr/Determ

Figure 17

Remote status diagram

EOD = End of Determination

10.7

Remote control via an RS-232 connection

The 870 KF Titrino plus can be remote controlled via an RS-232 connec-
tion. For this, a 6.2148.030 RS-232/USB Box is necessary. Connect the
RS-232/USB Box to the USB interface of the instrument.

The RS-232/USB Box has two connectors for RS-232 instruments. The
RS-232 remote control functions only through the connector RS-232/2.
The connector RS-232/1 is designed for connection with a balance. Con-
nect the RS-232/2 connector of the RS-232/USB Box to a serial interface
(COM interface marked with the symbol IOIOI) of a PC. Use a
6.2134.040 connection cable.