9 Troubleshooting

870 KF Titrino plus

■■■■■■■■

67

9 Troubleshooting

9.1

Karl Fischer titration

Problem

Cause

Remedy

The drift is very high
during conditioning.

The titration cell is leaking.

Check seals and septum. If necessary,
replace.

Replace molecular sieve.

The drift becomes
greater after each
titration.

The sample releases water
very slowly.

Adjust the method.

Add solubilizer.

Increase the temperature (possibly using a
KF oven).

See technical literature.

A side reaction is taking
place.

Use special reagents.

Adjust the method (increase/decrease the
temperature, external extraction).

See technical literature.

The pH value is not in the
optimum range anymore.

Add buffer, see technical literature.

The titration will not
be finished.

The titration cell is leaking.

Check seals and septum. If necessary,
replace.

Replace molecular sieve.

The minimum increment is
too low.

Define Titration rate = user and increase the
minimum volume increment (Min. incre-
ment
) (see Chapter 7.1.3, page 53).

The stop criterion is unsuit-
able.

Adjust the control parameters (see Chapter
7.1.3, page 53)
:

Increase the stop drift.

Select a short delay time.

See also: The drift increases
after each titration.