5 instrument diagnosis, 1 loading program versions and language files, Instrument diagnosis – Metrohm 870 KF Titrino plus Manual User Manual

Page 59: Loading program versions and language files

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 System settings

870 KF Titrino plus

■■■■■■■■

49

6.5

Instrument diagnosis

6.5.1

Loading program versions and language files

Menu

System Diagnosis

New program versions or language files can be loaded from a USB flash
drive. The corresponding file must be saved on the USB flash drive in a
directory with the instrument number (e.g. 848 or 863).

You can distinguish between language files and program files by noting
how the file name is constructed.

Program files
They are instrument-specific. The file name has the following structure:

5XXXyyyy.bin where

XXX =

yyyy =

Instrument type (e.g. 848 for the 848 Titrino plus)

Program version

Language files
They can be recognized by means of the two-digit language code in the
file name. A language file contains the dialog texts for various instrument
types. It is not instrument-specific. The file name has the following struc-
ture:

5848xxxxYY.bin where

xxxx =

YY =

Version number

Language, e.g. DE (German), FR (French), ES (Spanish)

Loading a file

Proceed as follows:

1 Connecting the USB flash drive

Plug in the USB flash drive with the 6.2151.100 adapter (USB
MINI (OTG) - USB A) at the instrument's USB port.

Switch on the instrument.

2 Opening the update dialog

Under Menu

System Diagnosis, select the menu item

Software update.

Press [OK].

Advertising