5 Operation

870 KF Titrino plus

■■■■■■■■

35

Continuous dosing (DOS)
Continuous dosing will be carried out with the DOS function for as long
as you keep the [START] key pressed down.

Proceed as follows:

1 Open the manual control

In the main dialog, select Menu and press [OK].
The main menu opens.

Select the menu item Manual control and press [OK].

Manual control opens.

2 Select the dosing function

Select the Dosing item.

In the function bar, select DOS and press [OK].

3 Configure the dosing function

NOTE

The dosing and filling rates should be decreased for viscous liq-
uids.

The maximum dosing and filling rates depend on the cylinder
volume (see Chapter 10.1.1, page 70).

Enter the dosing rate.

Enter the filling rate.