7 calculations - methods "titer ipol/upol, Calculations - methods "titer ipol/upol – Metrohm 870 KF Titrino plus Manual User Manual

Page 70

Advertising
background image

7.1 Karl Fischer Titrations (KFT)

■■■■■■■■■■■■■■■■■■■■■■

60

■■■■■■■■

870 KF Titrino plus

Result unit

The result unit is displayed and saved together with the result. It cannot
be modified.

Selection

mL

7.1.7

Calculations - Methods "Titer Ipol/Upol"

Menu

Parameters Calculation

Calculation formula

(C00*FCT)/EP1

The calculation formula is predefined and cannot be edited.

The calculated result (value and unit) is saved as the titer of the solution
used.

The Factor (FCT) parameter must be adjusted according to what is used
to determine the titer of the reagent and which unit the sample size has. If
the calculation formula is selected and the [OK] key is pressed, then a
table with the conversion factors will be displayed:

Table 2

Conversion table

Standard used

Sample size
in …

Factor (FCT)

Water Standard 10 mg/g

g

Water content in mg/g
(see certificate)

Water

g

1000

Water

µL

Density of water in
g/mL

Sodium tartrate dihydrate

g

156.6

Sodium tartrate dihydrate

mg

0.1566

Factor (FCT)

Conversion factor, see table above.

Input range

–999999999 - 9999999999

Default value

1.0

Decimal places

Number of decimal places used to display the result.

Input range

0 - 5

Default value

4