2 editing the solution data, Editing the solution data – Metrohm 870 KF Titrino plus Manual User Manual

Page 53

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 System settings

870 KF Titrino plus

■■■■■■■■

43

6.2.2

Editing the solution data

Name

The designation of the solution is used for unambiguous identification.

Entry

max. 24 characters

Default value

empty

Type

The model of the buret unit is displayed.

Cylinder volume

Cylinder volume of the buret unit in mL. The cylinder volume is automati-
cally read out with intelligent buret units.

Selection

1 | 5 | 10 | 20 | 50

Default value

20

Concentration

Concentration of the solution.

Input range

–999999999 - 9999999999

Default value

1.000

Concentration unit

Unit of the concentration.

Selection

µmol/mL | mmol/L | mol/L | g/L | mg/L | mg/mL |
µg/L | ppm | % | mEq/L | User-defined

Default value

mol/L

User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined. A blank entry can be generated this way as well.

Titer

Titer of the solution.

Input range

–999999999 - 9999999999

Default value

1.000

Titer unit

Unit of the titer.

Selection

µmol/mL | mmol/L | mol/L | g/L | mg/L | mg/mL |
µg/L | ppm | % | mEq/L | empty | User-defined

Default value

empty