7 Parameters

870 KF Titrino plus

■■■■■■■■

51

7 Parameters

7.1

Karl Fischer Titrations (KFT)

7.1.1

Conditioning

Menu

Parameters Conditioning

Under Conditioning, the conditions required for conditioning are
defined.

Conditioning

If this parameter is switched on, then the first time the titration is started
the working medium will be titrated to the endpoint with the specified
control parameters. The status is kept stable. The actual method run does
not begin until [START] has been pressed once more. Conditioning will
be carried out again automatically after the titration.

Selection

on | off

Default value

on

Start drift

Conditioning OK will be displayed as soon as this volume drift has been
reached and the titration can be started.

Input range

1 - 999 µL/min

Default value

20 µL/min

Drift correction

The endpoint volume can be corrected by taking the drift value into
account. For this, the volume drift is multiplied with the drift correction
time and this value is then subtracted from the endpoint volume. The drift
correction time is the time interval between the end of conditioning and
the end of the determination.

Selection

auto | manual | off

Default value

off

auto
The value of the current volume drift is applied automatically at the
start of the titration.
manual
If the volume drift is known throughout a prolonged period of time,
then this can be entered manually.
off
No drift correction takes place.