9 Troubleshooting

870 KF Titrino plus

■■■■■■■■

69

Problem

Cause

Remedy

The printer model is not
compatible.

Use a printer which fulfills the required specifi-
cations (see Chapter 10.4.4, page 74).

The USB keyboard
or the PC mouse
does not function.

The keyboard or mouse is
not recognized by the
instrument.

Switch the 870 KF Titrino plus off and on
again.

Use a hub on the USB connector and con-
nect the keyboard or mouse to the hub.

The keyboard or the mouse
is not compatible.

Use a model which fulfills the required specifi-
cations (see Chapter 10.4, page 72).

The display is no
longer readable.

The contrast is set incor-
rectly.

Adjust the contrast correctly (see Chapter 6.1,
page 39)
.

Mettler XP balances
send "R" or "O ----"
as ID1.

The automatic calibration
of the balance is switched
on.

Switch off the automatic calibration.

Message 020-511
"Action not possi-
ble" is displayed.

The USB stick is no longer
connected.

1. Connect the USB stick.
2. Switch the instrument off and on again.

The USB stick is full.

Use a different USB stick.

Delete files with the aid of a PC.