Metrohm 870 KF Titrino plus Manual User Manual

Page 72

Advertising
background image

7.1 Karl Fischer Titrations (KFT)

■■■■■■■■■■■■■■■■■■■■■■

62

■■■■■■■■

870 KF Titrino plus

Result unit

Sample size
in …

Factor (FCT)

Divisor (DIV)

mg/mL

mL

1

1

mg/piece

pieces

1

1

Factor (FCT)

Conversion factor, see table above.

Input range

–999999999 - 9999999999

Default value

0.1

Divisor (DIV)

Conversion factor, see table above.

Input range

–999999999 - 9999999999

Default value

1.0

Titer

Titer of the solution used. As soon as a solution has been selected under
Titration parameters, the value and the unit will be read out from the
solution data under System

Solutions and displayed here. If the titer is

changed here manually, then the solution data will be updated.

NOTE

Only the numerical value is taken into account in the calculation for-
mula. The calculation will then provide a correct result only if the unit
of the titer is mg/mL.

Input range

0.00000001 - 9999999999

Default value

1.000

Blank (CV01/2)

This parameter is visible only with the methods KFT Ipol–Blank and KFT
Upol–Blank
.

This value is calculated with the blank value determination (methods
Blank Ipol and Blank Upol) and entered here. The value can also be
modified manually.

Input range

–999999999 - 9999999999 mL

Default value

0.0 mL

Decimal places

Number of decimal places used to display the result.

Advertising