10 Appendix

870 KF Titrino plus

■■■■■■■■

75

Proceed as follows for the system initialization:

1 Switching off the instrument

Keep the red [STOP] key pressed down for at least 3 s.

A progress bar is displayed. If the key is released during this time,
then the instrument will not be switched off.

2 Switching on the instrument

Keep the red [STOP] key pressed down for approx. 10 s.

The dialog for confirmation of the initialization is displayed for 8 s.
The initialization must be confirmed during this time.

3 Confirming the initialization

NOTE

If the request is not confirmed within 8 s, then the procedure will
be canceled.

Press [BACK] twice.

Initialization is started. The process takes approximately 80 s. The
instrument will be automatically restarted after successful initializa-
tion.