7 Parameters

870 KF Titrino plus

■■■■■■■■

61

Result unit

The result unit is displayed and saved together with the result.

Selection

% | ppm | mg/mL | g | mg | mL | mg/piece |
User-defined

Default value

mg/mL

User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined. A blank entry can be generated this way as well.

7.1.8

Calculations - Methods "KFT Ipol/Upol", "KFT Ipol/Upol–Blank"

Menu

Parameters Calculation

Calculation formula for KFT Ipol/Upol

EP1*TITER*FCT/(C00*DIV)

Calculation formula for KFT Ipol–Blank

(EP1–CV01)*TITER*FCT/(C00*DIV)

Calculation formula for KFT Upol–Blank

(EP1–CV02)*TITER*FCT/(C00*DIV)

The calculation formulas are predefined and cannot be modified.

The Factor (FCT) and Divisor (DIV) parameters need to be adjusted,
depending on the unit used for the sample size and the result. If the calcu-
lation formula is selected and the [OK] key is pressed, then a table with
the conversion factors will be displayed:

Table 3

Conversion table

Result unit

Sample size
in …

Factor (FCT)

Divisor (DIV)

%

g

0.1

1

%

mg

100

1

%

mL

0.1

Density of the
sample in g/mL

ppm

g

1000

1

ppm

mL

1000

Density of the
sample in g/mL

mg/mL

g

Density of the
sample in g/mL

1

Advertising