Metrohm 774 Oven Sample Processor User Manual

Page 105

Advertising
background image

5.8 Remote Interface

774 Oven Sample Processor, Instructions for use

99

CTL-command

Parameter Bit

Pattern

Function


INIT

00000000000000

initializes the remote interface

START device1

*************1

starts device1 (for ex., Titrino, Coulometer...)

START dos1

*******1******

starts Dosimat on device1 (Titrino via "activate",

6.2141.040 cable required)

STOP device1

************1*

stops device1 (for ex., Titrino, Coulometer...)

START 737

*********1****

starts Coulometer 737 (6.2141.000 cable required)

ENTER

*********1111*

simulates the <ENTER> key of Titrinos or a

756 Coulometer


With the START and STOP commands a short pulse (200 ms) is emitted.


Manual stop options

Parameter Bit

Pattern

Function

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
STOP device1

************1*

stops device1 (for ex., Titrino, Coulometer...)



With the STOP command a short pulse (200 ms) is emitted.



SCN Command

Parameter Bit

Pattern

Function

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
Ready1

*******1

queries "ready" state of device1 (Titrinos. …)

End1

****1***

expects the end pulse of device1 (EOD)

Cond ok

******1*

queries "Cond" state of device1 (Titrinos. …)

Cond 737

1****1**

queries "Cond" state of the 737 Coulometer

no error

**1*****

queries "no error" state of device1 (Titrinos. …)