5 Detailed description

774 Oven Sample Processor, Instructions for use

70

parameters

>report

Submenu for report type definition

Open the submenu with <ENTER>

>report

report: none

Setting of the report type

none

fulll

short

config

param

usermeth

all

= No report output

= Output of the oven temperatures

= Short result report

= Listing of the device configuration

= Method listing

= List of all stored methods

= Printout of all above reports


When selecting a result report (full or short), a separate report is printed out
for each sequence (start, sample and final sequence).


>report

rec. interval

10 s

Time interval for recording the oven
temperatures

1…10…60 s

Maximum 400 temperature values (meas-

urement points) can be recorded per se-

quence. This results in a maximum re-

cording time of approx. 6…7 hours in the

case of one measuring report per minute.

The recording starts at the start of a se-

quence in each case.


parameters

>changer settings

Submenu for the changer settings

Open the submenu with <ENTER>

>changer settings

rack number

0

The rack that is assigned to the
method

0…16

0 = no particular rack

This setting can force the use of a certain rack with the method chosen. If this
is not desired, the rack number 0 must be chosen.

>changer settings

lift rate 1

12 mm/s

Stroke speed of lift 1

3…12 mm/s

>changer settings

lift rate 2

12 mm/s

Stroke speed of lift 2

3…12 mm/s

>changer settings

shift rate.

20

Turning speed of the rack in angular
degrees/second

3…20

>changer settings

shift direction:

auto.

Turning direction of the rack

+, -, auto.

auto. =

the sample changer chooses

the shortest path for turning.