5 Detailed description

774 Oven Sample Processor, Instructions for use

64

>auxiliaries

beaker sensor:

ON

Activate or deactivate beaker sensor

ON,OFF


configuration

>oven settings

Submenu of oven settings

Open submenu with <ENTER>

>oven settings

initial temp.:

OFF °C

Setting the initial temperature of the
oven after switching on

OFF, 50…250 °C

OFF = Do not switch on oven heater

>

oven settings

max.temperature 275

°C

Maximum possible temperature
(overheating safeguard)

50…275 °C

>oven settings

temp.correction 0

°C

Correction value for heater control

-10…0…10 °C


configuration

>rack definitions

Submenu for the definition of the individual racks

Open the submenu with <ENTER>

>rack definitions

rack number

1

Number of the rack

1…16


The number of the rack in position is automatically displayed here. If the con-
figuration of another rack must be changed, its rack number must be entered
and confirmed with <ENTER>. The rack number will be shown in the first
menu line for the subsequent entries.


>rack definitions

1

code 000001

Identification code of the rack

6 Bits

The rack code must be unique and can only occur once in the instrument.

>rack definitions

1

type :

M36-0

Type description of the rack
See p. 88ff.

M36-0


<SELECT> enables the choice of Metrohm-specific and self-defined rack
types.

Use

<QUIT>

to access next

higher level

Advertising