2 gas flow, Gas flow diagram, Display – Metrohm 774 Oven Sample Processor User Manual

Page 45

Advertising
background image

4.2 Gas flow

774 Oven Sample Processor, Instructions for use

39

Gas flow diagram:

inlet filter

flow meter

pump

pressure control

valve

gas flow

control knob

solenoid
valve

Air/N2 in

from Drying flask

to Drying flask

tower

outlet needle

injection

needle

sample

titration cell

conveying tubing



left-hand

instrument

side

Gas flow

Air/N

2

in

The gas flow can be set manually with the help of the gas flow regulator and
the flowmeter. The flow rate of the carrier gas is shown on the display in
mL/min or L/h so that it is easy to set the required gas flow with the rotary
knob of the gas flow regulator. A gas flow of 20 to 100 mL/min is required;
this depends on the determination method and moisture in the sample.

Display:

******** counter 1/36

110°C 55 mL/min ready

or:

(

press <DISPLAY>)

gas flow 55 mL/min

heater temp. 120°C


The gas flow is monitored automatically. A lower and an upper limit can be
defined for the gas flow; if these are undercut or exceeded an error mes-
sage is displayed. It is recommended that the safety limits (min. flow and
max. flow) are not set too closely for gas flow monitoring. A range of ap-
prox. 30 mL/min should be set.

During needle penetration and removal the gas flow should be switched off
as large variations in gas pressure occur during these processes. This
avoids unnecessary error messages.