Metrohm 774 Oven Sample Processor User Manual

Page 24

Advertising
background image

2 Installation

774 Oven Sample Processor, Instructions for use

18

Control commands:

CTL:Rm : START Gerät1

starts

Titrino

CTL:Rm : ***********1**

advance impulse / ENTER


Scanning via remote lines:

SCN:Rm : ****1000

waits for end of titration (EOD-impulse)

SCN:Rm : *****010

waits for 'cond. ok'

In principle it is possible to use an additional RS232 connection to auto-

matically load a particular method in the Titrino while a sequence is being

processed. Please refer to the example of the remote connection given for

the 756 Coulometer on the previous page.


774 Oven Sample Processor — 796 Titroprocessor
for volumetric KF titrations


If connected to a 774 Oven Sample Processor the 796 Titroprocessor takes
over the functions of the control instrument itself (Master). The communica-
tion between the 774 Oven Sample Processor and the 796 Titroprocessor
can be solved in any manner.

If a remote connection is used (6.2141.020 cable) the input and output lines
0…7 can be used as required. However, it is then not possible to start a
774 Oven Sample Processor method. See page 79ff for the communication
commands.

A serial RS232 connection between the 796 Titroprocessor and the 774
Oven Sample Processor has the advantage that the Titroprocessor as con-
trol unit can access all internal functions except starting or loading a
method in the 774 Oven Sample Processor. Please refer to the description
of the Metrohm remote language and the 774 remote control tree on page
98ff.

6.2141.020

cable (remote)

6.2125.110

cable (RS232)

or

Advertising