5.2 The keyboard

774 Oven Sample Processor, Instructions for use

51

Key

Normal state

Editing

FLOW

4

Activating/deactivating gas
flow

• Key <FLOW> is used to

activate resp. deactivate the
air pump or the solenoid valve
(Air/N

2

in).

• The first parameter serves to

select pump or valve. The
second parameter defines
status 'On/Off'. The function
must be confirmed with
<ENTER>.

Numeric entry ('4')

or

activate/deactivate gas flow

• The FLOW command serves to

activate resp. deactivate the air
pump or the solenoid valve
(Air/N

2

in).

• The first parameter serves to

select pump or valve. The sec-
ond parameter defines status
'On/Off'.

5

HEATER

Oven control
• Key <HEATER> serves to

set the current oven tempera-
ture.

• The first parameter defines the

oven temperature. With the
second parameter, it is possi-
ble to achieve a slow, time-
controlled heating-up process
to the set target temperature.
You can set a time interval
within which the target tem-
perature should be reached.
The function must be con-
firmed with <ENTER>.

Numeric entry ('5')

or

oven control
• The first parameter defines the

oven temperature (target tem-
perature).

• Using the second parameter, it

is possible to program tem-
perature ramps or (by combin-
ing several HEATER com-
mands) entire temperature pro-
files. The heat-up interval de-
fines the period within which
the set target temperature
should be reached.

DOS

6

Dosing control
• The <DOS> key serves for

the control of Dosimats and
Dosinos. These are controlled
via the "External Bus" connec-
tor.

• The first parameter is for the

selection of the dosing unit.
The second parameter repre-
sents the function.

• In addition to dosing volume

and the fill command, the
functions specific to the Dos-
ino -’prepare’, ‘release,
‘empty’ and 'release'- can also
be executed.

• The dosing and filling speed

can be set in the parameter
menu or with the <DEF> key.

Numerical entry ('6')

or

Dosing control
• In a sequence you can not only

add a certain volume under
process control, it is also pos-
sible to initiate the burette filling
with Dosimats or initiate the
designated functions with Dos-
inos as there are: filling, empty-
ing, preparation or release of
the exchange unit.