5 initialise data memory – Metrohm 774 Oven Sample Processor User Manual

Page 155

Advertising
background image

6.5 Initialise data memory

774 Oven Sample Processor, Instructions for use

149

Magnets

fitted

In this

example, the code is:

000110

Magnet fixture viewed from below

• Raise the rack and place it to one side.

• Prepare the unit for diagnosis (see Chap. 6.4.2).

• If necessary, press<Ð> several times until

diagnosis

>power on reset

• <ENTER>

The unit runs through the power-up routine (initialisation of lift and rack positions). Initialisation is

important since it is essential that the turntable (rack holder) be in basic position (beaker 1 at

tower 1) for the following “rack code test“ diagnosis step.

• Prepare the unit for diagnosis (see Chap. 6.4.2).

• If necessary, press <Ð> several times until

diagnosis

>rack code test

• <ENTER>

The test reads in the coding constantly and displays it. A six-digit bit pattern (code ??????) is

scheduled for this for the display. The first digit for magnet No. 1, the second digit for magnet

No. 2 etc. If a magnet is detected, the corresponding digit is assigned "1". Otherwise, it is as-

signed "0".

• Consecutively fit all racks to be used and compare the related note

which you made beforehand with the information on the display.

>rack code test

code 000000

Example of:

No rack fitted

>rack code test

code 000001

Example of:

Rack with coding in accordance with above

example.

• You can quit the test by pressing key <QUIT> or <STOP>.

diagnosis

>power on reset


You can quit diagnosis from the main menu with <QUIT> or <STOP>.

6.5 Initialise data memory

This diagnosis step allows you to assign default values to device parame-
ters using the keypad and thus set the unit to its original state. This is im-
portant owing to the following two aspects: