5.3 Menu organization

774 Oven Sample Processor, Instructions for use

69

parameters

>start sequence

Editor for the start sequence of a sample series

Open the submenu with <ENTER>

The sequence entered here is run once when a sample series is started.
This may be used to process blind samples for instance.

parameters

>sample sequence

Editor for processing sequence for each sample

Open the submenu with <ENTER>

This sequence is run when processing each individual sample
of a series.

parameters

>final sequence

Editor for the end sequence of a sample series

Open the submenu with <ENTER>

This sequence is run once at the end of a sample series. This may be re-
setting of the sample rack for instance.


In principle the same rules for entry are valid here as for manual operation,
i.e. after selecting a command and entering the necessary data, the entry is
accepted with <ENTER>. Afterwards the next command line is accessed
where a new command can be entered.

The LEARN mode is available for the easy entry of parameters. With certain
commands “live” values can be taken over by manual execution of a single
command. More information can be found on page 33.

Furthermore the TRACE function can be used to execute every command
line step by step. See page 34.

Navigation in a sequence is accomplished as in the other menus. In addi-
tion the <INSERT> and <DELETE> keys can be used.

<INSERT> adds a new command line above the current line in a se-
quence. It is automatically occupied by the “NOP” command that has no
function. The following lines are shifted one line downwards.

<DELETE> deletes the current line in a sequence. The lines following are
shifted one line upwards.