5 Detailed description

774 Oven Sample Processor, Instructions for use

68

PARAM

5.3.2 Parameters


All settings of the Parameter Menu constitute a method and may be saved as such.

Main Menu:

parameters

number of samples

rack

Number of samples to be processed

1…999,

rack,*

rack = one entire rotation of the rack
*

= infinite number of samples

All sample positions of an engaged rack will be processed when the instrument is on the ‘rack’
setting. (max. number of rack positions – number of special beakers defined). It is important that
the changer can recognize the rack. This is only possible when the rack is at the ground position.
It is recommended to initialize the changer with the <RACK> key.

parameters

>start sequence

Open the submenu with <ENTER>

parameters

>sample sequence

Use <

Ï

> or <Ð> to move up or down one menu item

parameters

>final sequence

Use <HOME> or <END> to move to first or last menu
item respectively

parameters

>report

Return to the normal state with <QUIT>

parameters

>changer settings


parameters

>timeout settings


parameters

>gas flow

parameters

>dosing unit def.


parameters

>manual stop




Submenus:

Up to 99 lines can be entered as a sequence in submenus '>Start Sequence', '>Sample Se-
quence' and '>Final Sequence'. The commands can be entered directly on the keyboard. The
command keys arranged at the right-hand half of the keyboard are available.