5.11.3.14

Info.ActualInfo.Inputs.Status read

only

Status of the Input lines (Input0…7) of the Remote interface.
$Q sends the signal state as decimal number e.g.
10 ⇒ 00001010 binary ⇒ 2

1

+ 2

3

⇒ Input1 and Input3 active

(active = low, inactive = high)
See also page 121ff.

5.11.3.15

Info.ActualInfo.Outputs.Status read

only

Status of the output lines (Output0…13) of Remote interface. See 5.11.3.14.

5.11.3.16

Info.ActualInfo.Display.L1 read

only

Info.ActualInfo.Display.L2 read

only

Text of the first and second line of the LCD display.

5.11.3.17

Info.ActualInfo.Counter.Sample read

only

Info.ActualInfo.Counter.Maximum read

only

Currently processed sample number and total number of samples.