Metrohm 774 Oven Sample Processor User Manual

Page 23

Advertising
background image

2.5

Integration

774 Oven Sample Processor, Instructions for use

17

Control commands of the 774:

CTL:Rm : START Gerät1

starts

Coulometer

CTL:Rm : *************1

"

CTL:RS &U.R.N

loads a method in the Coulometer

CTL:RS "774BLANK"..$G

here e.g. "774BLANK"


Scanning the remote lines of the 774:

SCN:Rm : **0**010

waits for 'cond. ready'

SCN:Rm : *****000

waits for end of determination


774 Oven Sample Processor — 737 Coulometer
for coulometric determinations

774

737



The 737 Coulometer is completely controlled via the remote lines.

Control commands:

CTL:Rm: *********1****

starts Coulometer with

CTL:Rm: *********0****

a start impulse

Scanning via remote lines:

SCN:Rm : 10000100

waits for 'cond. ready'


774 Oven Sample Processor — 7xx Titrino
for volumetric KF titrations

774

758

6.2141.000 cable (remote)

6.2141.020

cable (remote)