Metrohm 774 Oven Sample Processor User Manual

Page 123

Advertising
background image

5.11 Description of the remote control commands

774 Oven Sample Processor, Instructions for use

117

5.11.1.7

Mode.SampleSeq.1.

.Move…, .Lift…, .Sample…,

etc. up to .100

.Heater…,.Def…, .Flow…, Dos…,

Scan…, .Ctrl…, .Wait…, .Rack, .Nop

Indexed sample sequence; its commands will be executed line by line in
processing. See start sequence (5.11.1.5).

5.11.1.8

Mode.FinalSeq.1.Cmd

NOP, MOVE, LIFT, STIR,

etc. up to .100

DEF, PUMP, DOS, SCAN,

SAMPLE, CTRL, WAIT, ENDSEQ

Defines the command of the indexed command line in a final sequence.
See start sequence (5.11.1.4).

5.11.1.9

Mode.FinalSeq.1.

.Move…, .Lift…, .Sample…,

etc. up to .100

.Heater…,.Def…, .Flow…, Dos…,

Scan…, .Ctrl…, .Wait…, .Rack, .Nop

Indexed final sequence; its commands will be executed line by line in proc-
essing. See start sequence (5.11.1.5).

5.11.1.10

Mode.Report.Assign

none, short, full, param,

config,

Usermeth

Mode.Report.TDelta 1…10…60
Report definition.
Only one report is possible. In the case of a 'full' result report, a measuring
point list with the oven temperatures as a function of time is output by con-
trast with the 'short' result report. The time interval (in seconds) between the
measuring points defines the node .TDelta. 400 measuring points are pos-
sible, see 5.11.3.2.

5.11.1.11

Mode.Changer.RackNo

0…16

Mode.Changer.L1Rate 3...12 mm/s

Mode.Changer.L2Rate 3...12 mm/s

Mode.Changer.ShRate 3...20 w/s

Mode.Changer.ShDir +,

–,

auto.

Mode.Changer.ModeSample

MOVE, display

Changer settings.
RackNo:

Rack number, forces the use of the defined rack with the
current method (0 = any rack).

L1Rate:

Lift speed for tower 1, in mm/sec

L2Rate:

Lift speed for tower 2, in mm/sec

ShRate:

Turning speed of the rack, in angular degrees/sec

ShDir:

Turning direction of the rack (ascending or descending
rack positions; auto. means automatic choice of the
shortest path)

ModeSample:

Reaction on missing sample beaker. (MOVE = next sam-
ple beaker will be chosen regarding the recent SAMPLE
command, display = display warning.)

5.11.1.12

Mode.TimeoutSet.HTime

off, 0...999

Mode.TimeoutSet.HAction cont,

STOP

Mode.TimeoutSet.STime

off, 0...999

Mode.TimeoutSet.SAction cont,

ERROR

Timeout settings. Timeout times are defined in seconds.
The HEATER timeout occurs if the set oven temperature (in the case of the
HEATER command) is not reached after the specified heating (in minutes).