6.5 Initialise data memory

774 Oven Sample Processor, Instructions for use

151

• Press key <ENTER> to open the following Diagnosis menu:

>RAM initialization
select: param

9

>RAM initialization
select: config

9

>RAM initialization
select: setup

9

>RAM initialization
select: assembly

9

>RAM initialization
select: all

9

9

9

9

9

Set method parameters to default val-

ues.

©

Set device configuration to default

values.

©

Set Setup parameters to default values.

©

Set Assembly parameters to default values.

©

Erases all user-defined methods

©

The sub-menus can be selected one after the other by pressing key <Se-
lect>. You can access the individual initialisation variants with key
<ENTER>, and you can quit with key <QUIT>.

The table shows what parts of the data memory are affected by the corre-
sponding initialisation variants. In the case of a system crash (undefined
display), no responses to keystrokes etc.), it is advisable to use initialisation
variant "

all

".

• If necessary, press key <Select> several times until:

>RAM initialization

select: all

• Press <ENTER>.

diagnosis

>RAM test

• Press <QUIT>.

The Diagnosis menu is closed and the device runs through a power-up reset.

Advertising