Metrohm 774 Oven Sample Processor User Manual

Page 88

Advertising
background image

5 Detailed description

774 Oven Sample Processor, Instructions for use

82

This function is suitable for instruments with Metrohm remote control lan-
guage. These can be controlled with so-called triggers.
The most important of these are:

&M;$G

Go, start instrument in current mode

&M;$S

Stop, stop instrument

&M;$H

Hold, interrupt method

&M;$C

Continue, resume method


The following remote control commands can switch on the AutoInfo status
messages of a Metrohm instrument (for example, in a start sequence):

&Se.A.T.R"ON" 'Ready' status message
&Se.A.T.F"ON"

status message at the end of a determination

&Se.A.T.S"ON" status message at manual halt
&Se.A.T.G"ON" status message at the start of a method
&Se.A.T.E"ON" status message during an error condition


To be consistent, the corresponding AutoInfo messages should also be
switched off again in a final sequence (..."OFF").

Detailed information about the syntax of the Metrohm remote control lan-
guage can be found in Chapter “Operation via the RS 232 interface" (page
99) or in the instruction manual of your instrument.

Please keep the syntax and conventions of the foreign instruments or com-
puters the Sample Changer is communicating with.



WAIT

3

WAIT

>sample sequence

11 WAIT

1 s

0…1…9999 s

Waiting time

The WAIT command sets a particular waiting interval during method proc-
essing.


RACK

>

*

RACK

-

>final sequence

1 RACK

Reset sample rack

The RACK command serves to reset the sample rack. The lift is moved to
initial position and the rack is turned to the basic position so that it can be
removed.


Advertising