7 rs232 interface, 8 external bus interface, Chap. 6.4.7 – Metrohm 774 Oven Sample Processor User Manual

Page 153: Chap. 6.4.8

Advertising
background image

6.4

Diagnosis

774 Oven Sample Processor, Instructions for use

147

diagnosis

>remote test

• <ENTER>

>remote test

remote test connector ?

• Connect the test connector 3.496.8550 to the Remote socket without

switching the unit off.

• <ENTER>

The test runs automatically. If no error occurs, the following is displayed:

>remote test

remote test ok

• Disconnect the test connector and press <ENTER>.

6.4.7 RS232 interface

This diagnosis step allows you to check all outputs and inputs for operabil-
ity.

• Prepare the unit for diagnosis (see Chap. 6.4.2).

• If necessary, press <Ð> several times until

diagnosis

>RS232 test

• <ENTER>

>RS232 test

RS232 test connector ?

• Connect the test connector 3.496.8480 to socket RS232 without switch-

ing the unit off.

• <ENTER>

The test runs automatically. If no error occurs, the following is displayed.

>RS232 test

RS232 test ok

• Disconnect the test connector and press <ENTER>.


6.4.8 External bus interface

This diagnosis step allows you to check all parts of the external bus inter-
face for operability.

• Prepare the unit for diagnosis (see Chap. 6.4.2).