Metrohm 774 Oven Sample Processor User Manual

Page 124

Advertising
background image

5 Detailed description

774 Oven Sample Processor, Instructions for use

118

HTime is the additionally granted waiting time for still reaching the target
temperature. HAction is executed after HTime elapses.
The SCAN timeout is triggered on execution of a SCAN command. STime
determines the reliable response time of the connected device (generally
the determination time). SAction is executed after expiry of STime.

cont:

Continue without action

STOP: Stop

method

ERROR:

Trigger error message

5.11.1.13

Mode.Gas.UnitFlow

mL/min, L/h

Mode.Gas.MinFlow

0…999

Mode.Gas.MaxFlow 0…900…999

Mode.Gas.Select

Air, others, N2

Mode.Gas.Otherfac 0.001…1…9.999
.UnitFlow determines the unit of the gas flow rate. The two nodes .Minflow
and .Maxflow define the limits which, if undershot or overshot, trigger error
messages (RS error E163 resp. E206).
Node .Otherfac takes effect only if 'others' is selected at the .Select node.

5.11.1.14

Mode.DosimatSet.DosUnitNo

1…12

Mode.DosimatSet.1.DosRate 0.01…160

ml/min,

max.

Mode.DosimatSet.1.FillRate 0.01…160

ml/min,

max.

Mode.DosimatSet.1.DosTube

1...4

Mode.DosimatSet.1.FillTube 1...2...4

Mode.DosimatSet.1.ExchTube 1...2...4

Mode.DosimatSet.1.PrepTube

1...4

Mode.DosimatSet.1.EmptyTube 1...4

etc.

until

.12

Dosing unit settings.
DosUnitNo:

Number of current dosing unit

DosRate: Dosing

speed

FillRate: Filling

speed

DosTube:

Dosing outlet of 700 Dosino

FillTube:

Filling inlet of 700 Dosino

ExchTube:

Rinsing inlet on exchanging the exchange unit (see
&Assembly.Dos.Value("release") 5.11.6.6)

PrepTube:

Dosing outlet on preparing cycle of 700 Dosino

EmptyTube:

Air inlet for emptying the 700 Dosino.

5.11.1.15

Mode.ManStop.RemCtl

STOP device1, 14 Bit (1,0, or ∗)

Mode.ManStop.RSCtl

&M;$S, 14 ASCII characters

Mode.ManStop.Flow

cont., off

Mode.ManStop.HEATER

INIT, off, 50…250

Functions executed when the <STOP> key is pressed.

.RemCtl

Bit pattern defines the line statuses of the Remote socket

.RSCtl

Character string is transferred via the RS232 cable

.Flow

The pump and the solenoid valve remain in the

current state or are both deactivated

.HEATER

Oven control is either set to the initial temperature

or switched off. Any temperature may also be

selected