5.4 Command reference

774 Oven Sample Processor, Instructions for use

83

DEF

0

DEF

Redefining specific instrument settings

During method processing the most diverse settings can be made using
the following DEF commands. The individual entries are accessed by
repeatedly pressing the DEF key (roll-up selection).


>sample sequence

14 DOSRATE

1

160

1…12

0.01…160 ml/min

Dosing speed

The dosing speed for all 12 dosing units can be individually set. The first
parameter stands for the number of the dosing unit and the second pa-
rameter allows setting the dosing speed in ml/min.

>sample sequence

12 FILLRATE 1

160

1…12

0.01…160 ml/min

Filling speed

The filling speed can be set for all 12 dosing units individually. The first pa-
rameter stands for the number of the dosing unit and the second parameter
allows setting the filling speed in ml/min.

>sample sequence

16 LIFTRATE

1 12 mm/s

1 3…12 mm/s

Lift rate

The first parameter of this command defines the lift

on which the command is to be executed. Since

the 774 Oven Sample Processor is equipped only

with Lift 1, it is not practical to enter any other

value at this point.

>sample sequence

17 SHIFTRATE: auto. 20
auto.,+,– 3…20 w/s

Turning direction and speed


Turning direction and speed of the sample rack can be changed as de-
sired. The first parameter determines the turning direction.

auto. :

The changer automatically determines the shortest path.

+ :

The sample rack turns in the counterclockwise direction (to a
higher rack position)

- :

The sample rack turns in the clockwise direction (to a lower
rack position)


The second parameter determines the turning speed in angular de-
grees/sec.