Calibrating counter inputs (if present), Scaling event and daily total inputs (if present) – ADS Environmental Services Spider QR 775010 A0 User Manual

Page 44

Advertising
background image

ADS Environmental Services ADS Spider User Manual


Page 42

Calibrating Counter Inputs (if present)

You can calibrate Counter Inputs in two ways:

1. Scaling
2. Pre-setting the counter to a particular reading


Scaling is useful to make the counters read engineering values. For
example, the water meter may be producing one pulse per 5 gallons
of water flow. Hence, to make the counter register in gallons, you
would set the Scaler to 5.

NOTE: A Spider counts internally from 0 to 99,999 before it

wraps

around

to 0. The actual range will be multiplied by the

Scaler

. So, a

Scaler of

5

will produce an effective counter range of 0 to 499,995.


It is sometimes required to pre-set the counter to a particular value.
For example, you may wish to connect a water meter to a Spider and
make the Spider

s starting point the same as the reading on the

meter. To manually set the Spider

s starting point; simply enter the

required reading in the

Offset

field.

Scaling Event and Daily Total Inputs (if present)


You can scale Counter and Event Inputs by specifying a

Scaler

. A

new scaled value obtained by multiplying the raw counter reading by
the

Scaler

will be used for all functions, including logging and

alarms.

If scaling is not required, simply leave it set to

1

.


To change scaling for an input, first access the

CHANGE SETUP

page, as described in Setting up a Spider for the First Time on page
32. Select "Scale <your input>

from

Select Input to Calibrate or

Scale

. Enter the new

Scaler

value on the newly displayed page.


You can exit at any time without making changes by clicking the

Cancel

link. Save the new values by clicking on the

Save

button.

Advertising