3.3 Configuration Using the Front Panel (Legal for Trade)

The 480 indicator can be configured using a series of menus accessed through the indicator front panel when the
indicator is in configuration mode.

When the indicator is placed in configuration mode,

CONFIG

is shown on the display. The CONFIG menu is the

first of eight main menus used to configure the indicator. Detailed descriptions of these menus are given in
Section 3.3.1 on page 27. When configuration is complete, return to

CONFIG

and press the (

ZERO

) key to exit

configuration mode and save changes. Then replace the configuration switch access screw.

To place the indicator into the configuration mode, a configuration switch is accessed by removing the large
fillister head screw on the enclosure back. Switch is activated by inserting a nonconductive tool into the access hole
after pressing the menu switch.

1. Remove setup switch access screw from back of indicator.
2. Using a nonconductive tool, press the button located in the access hole. This allows access to configuration

menus.

Menu

Required

Password*

Menu Function

CONFIG

Configuration

Config

Graduations (scale capacity), Configure zero tracking, zero range, motion
band, overload, tare function, sample rate, and digital filtering parameters.

FORMAT

Format

Config

Set primary and secondary units.

CALIBR

Calibration

Config

Calibrate indicator. See Section 4.0 on page 37 for calibration procedures.

PROGRM

Program

Config

Set power-up mode, regulatory mode, and consecutive number values.

DIG IN

Digital Input

User

Assign digital input functions. Used only if the Relay option is installed.

ALGOUT

Analog Output

Config

Configure analog output module. Used only if analog output option is
installed.

PASWRD

Password

Config

Enable and edit configuration and user passwords.

TEST

Test

Config

Test A/D, Test and set Digital I/O, reset indicator to default settings, set over
and under weighment characteristics.

MENU

Menu

User

Non Legal-for-Trade settings, such as Audit Trail, view Tare, set Unit ID, view/
print Accumulator,set Time/Date, and Non Legal for Trade setting for Serial
Port, Setpoints, and Print Format parameters. See Section 3.2.

Can be accessed without pressing the Setup Switch by
pressing the MENU key on the front panel.

* Password required for all menus if a password for that level has been entered. See Section 3.3.7.

Table 3-6.

480

Menu Summary