4 more about calibration, 1 adjusting final calibration, Adjusting final calibration – Rice Lake 480 Legend Series Digital Weight Indicator Installation Manual User Manual

Page 43

Advertising
background image

Calibration 39

4.4 More About Calibration

The following provides additional information about compensating for environmental factors.

4.4.1

Adjusting Final Calibration

Calibration may be affected by environmental factors including wind, vibration, and angular loading. For example,
if the scale is calibrated with 1000 lb, a strain test may determine that at 2000 lb the calibration is 3 lb high. In this
case, final calibration can be adjusted by tweaking WVAL to 998.5 lb. This adjustment provides a linear correction
of 1.5 lb per 1000 lb.

To adjust the final calibration, return to the

WVAL

prompt and press

to show the test weight value. Press or

to adjust calibration up or down. Press

to save the value. Press to return to the CALIBR menu.