Calibration 37

4.0 Calibration

The 480 can be calibrated using the front panel or the Revolution

®

configuration utility. Each method consists of

the following steps:

• Zero calibration
• Entering the test weight value
• Span calibration
• Optional rezero calibration for test weights using hooks or chains.

The following sections describe the calibration procedure for each of the calibration methods.

4.1 Front Panel Calibration

The CALIBR submenu is used to calibrate the 480. See Section 3.3.3 for the calibration menu structure and
parameter descriptions. To calibrate the indicator using the front panel:

1. Remove the large fillister head screw from the back of the enclosure.
2. Insert a nonconductive tool into the access hole and press the configuration switch. Indicator display changes

to show

CONFIG

(see Figure 2-1).

3. Press

until

CALIBR

is displayed.

4. Remove all weight from the scale platform. If the test weights require hooks or chains, place the hooks or

chains on the scale for zero calibration.

5. Press

to

WZERO

.

6. Press

to

CAL

.

7. Press

to

ENTER

.

8. Press

to calibrate zero. The indicator displays

*CAL*

while calibration is in progress. When complete, the

mV value for the zero calibration is displayed.

9. Press to save the zero calibration value and go to

CAL

.

10. Press

to

WZERO

.

11. Press

to

WVAL

.

12. Press to show the test weight value.

13. Press to edit value using the procedure shown in Figure 3-3 to enter the actual value of the test weights to

be used for the span calibration. Press

to store the entered

WVAL

value.

14. Press

to

WVAL

.

15. Press

to

WSPAN

.

16. Press

to

CAL

.

17. Press

to

ENTER

.

18. Place test weights on the scale and press

to calibrate span. The indicator displays

*CAL*

while calibration

is in progress. When complete, the net millivolt value of the test weight for the span calibration is displayed.

Net mV value is equal to the raw mV value at the span minus the mV value captured for zero.

19. Press to save value and go to

CAL

.

20. Press

to

WSPAN

. If using chains or hooks for your test weights, proceed with step 21.

If no chains or hooks were used, proceed with step 23.

21. Press

to

REZERO

. The REZERO function is used to remove a calibration offset when hooks or chains are

used to hang the test weights.

22. Remove the hooks or chains used during calibration and remove the test weights from the scale. Press .

ZERO

will flash. With all weight removed, press

to rezero the scale. This function adjusts the zero and

span calibration values. The indicator displays

*CAL*

while the zero and span calibrations are adjusted.

23. Press to return to the CALIBR menu.

24. Remove all weight from the scale. Press until the display reads

CONFIG

. Press to exit configuration

mode.

The IN IZR (initial zero) parameter may affect the weight display reading immediately following calibration.

Note

Note