34

480 Operator’s Manual

3.3.6

Analog Output Menu

If the analog output option is installed, configure all other indicator functions and calibrate the indicator (see
Section 4.0 on page 37) before configuring the analog output. See Section 9.10 on page 59 for analog output
calibration procedures.

Figure 3-15. Analog Output Menu

ALG OUT Menu

Parameter

Choices

Description

SOURCE

GROSS
NET

Specifies the weigh data, gross or net, tracked by the analog output.

OFFSET

0
20

Zero offset in %. Select 0% for 0–10 V / 0-20 mA; select 20% for 2-10 V / 4-20 mA output.

ERRACT

FULLSC

Error action. Specifies how the analog output responds to system error conditions.

Set to full value (10 V or 20 mA)

HOLD

Hold current value

ZEROSC

Set to zero value (0 V or 0 mA) or (2 V or 4 mA)

MIN

000000
0 – 999999

Specifies the minimum weight value tracked by the analog output. Specify a weight value (in
primary units) in the range 0999999.

MAX

000010
0 – 999999

Specifies the maximum weight value tracked by the analog output. Specify a weight value (in
primary units) in the range 0999999.

MINNEG

NO
Yes

YES implies that the value in MIN is a negative value. NO is the default and implies that MIN is
a positive value.

MAXNEG

NO
Yes

YES implies that the value in MAX is a negative value. NO is the default and implies that MAX is
a positive value.

TWZERO

000000
065535

Tweak zero. Adjust the analog output zero calibration. Use a multimeter to monitor the analog
output value. Press

or

to select the digit to edit.

Press

or

to edit the selected

(flashing) digit. Press

to save the new value.

TWSPAN

000000
065535

Tweak span. Adjust the analog output span calibration. Use a multimeter to monitor the analog
output value. Press

or

to select the digit to edit.

Press

or

to edit the selected

(flashing) digit. Press

to save the new value.

OUTPUT

CURENT

Defines the output type.

0-20 mA or 4-20 mA Depends on OFFSET

VOLTAG

0-10 V or 2-10 V Depends on OFFSET

Table 3-12. Analog Output Menu Parameters

CONFIG

FORMAT

CALIBR

PROGRM

DIG IN

ALGOUT

PASWRD

TEST

MENU

SOURCE

NO

Yes

NO

YES

OFFSET

ERRACT

MIN

MAX

MINNEG

TWZERO

TWSPAN

MAXNEG

OUTPUT

000000

000000

CURENT

VOLTAG

GROSS

NET

0%

20%

000000

000010

FULLSC

HOLD

ZEROSC