9.1

Error Messages

The 480 indicator provides a number of error messages. When an error occurs, the message is shown on the
indicator LED display.

9.1.1

Displayed Error Messages

The 480 provides a number of front panel error messages to assist in problem diagnosis. Table 9-1 lists these
messages and their meanings.

Shorting the excitation voltage shuts the excitation voltage off. The only way to restore excitation voltage is to
cycle power.

Error Message

Description

Solution

E A/D

A/D physical error

Call Rice Lake Weighing Systems (RLWS) Service at 800-472-6703.

EEEROM

EEPROM physical error

EVIREE

Virgin EEPROM

Use TEST menu to perform DEFLT (restore defaults) procedure, then
recalibrate load cells.

EPCKSM

Parameter checksum error

EACKSM

A/D calibration checksum error

A/D converter requires recalibration. Call RLWS Service.

EFCKSM

Printer format checksum error

Call RLWS Service at 800-472-6703.

ELCKSM

Load cell calibration checksum error

Recalibrate load cells.

EIDATA

Internal RAM checksum error

Call RLWS Service at 800-472-6703.

E REF

A/D reference error

A/D converter requires recalibration. Call RLWS Service.

ERROR

Internal program error

Check configuration.
Call RLWS Service if unable to clear error by cycling power or if error
recurs.

OVERFL

Overflow error

Weight value too large to be displayed.

- - - - - -

- - - - - -

- - - - - -

Gross > overload limit

Gross value exceeds overload limit. Check configuration or signal
input level. Overload can be caused by input signal > 45 mV or
common mode voltage > 950 mV.

- - - - - -

Gross < 20d behind zero

Gross value is more than 20 divisions behind zero.

RNGERR

GRADS > 100,000
WVAL > 100,000

Only shows up in Config mode.

EEPERR

EEPROM error

Call RLWS for service at 800-472-6703.

HINOFF?

High offset

Zero load at powerup is more than initial zero range (INIZR) setting of
calibration zeroremove the extra load.

LINOFF

Low offset

Zero load at power up is less than initial zero range (INIZR) setting of
calibration zeroadd the missing load.

NOBATT

No battery

The RTC lost time/date tracking at previous power off state due to
low battery or no battery condition. The printer, accumulator and
AUDUT functions will fail to get time and date.

EUCKSM

Configuration checksum

The checksum value of configuration has changed from that stored
in memory.

OIMLER

OIML parameter error

Parameter set incorrectly for use in the OIML mode. Example:
Primary units set for lb or oz.

EE-ACC

Accumulator error

Error with the accumulator such as attempting to display an
accumulated value greater than six digits.