Configuration 27

3.3.1

Configuration Menu Structures and Parameter Descriptions

Figure 3-10. Configuration Menu

CONFIG Menu

Parameter

Choices

Description

GRADS

10000
1 – 100000

Graduations. Specifies the number of full scale graduations. The value entered must be in
the range 1 – 100 000 and should be consistent with legal requirements and
environmental limits on system resolution.

ZTRKBN

0 dd
0.0 – 3.0dd

Zero track band. Automatically zeroes the scale when within the range specified, as long
as the input is within the configured zero range (ZRANGE parameter). Selections are ±
display divisions. Maximum legal value varies depending on local regulations.

ZRANGE

2%
1% – 100%

Zero range. Selects the range within which the scale can be zeroed. The 2% selection is ±
2% around the calibrated zero point, for a total range of 4%. Indicator must be at
standstill to zero the scale.

INIZR

0 %
0 – 100%

Initial ZERO Range at power up. See Error messages HINOFF and LINOFF in
Section 9.1.1

If on power up, the weight value is between the ±% range specified of Calibrated Zero,
the indicator will zero off that weight. If outside that range, the indicator will display
HINOFF or LINOFF, depending on whether the initial reading is above or below the range.

If not set at zero, hopper/scale MUST be empty when powered up.

MOTBAN

0.1D
0.1D – 100D

Motion band. Sets the level, in display divisions, at which scale motion is detected. If
motion is not detected, the standstill symbol lights.

• Motion window is the consecutive A/D updates compared to the RD-MOT settings.
• Stability window is the consecutive A/D updates compared to the RD-STA settings.

Some operations, including print, tare, and zero, require the scale to be at standstill.
Maximum legal value varies depending on local regulations.

RD-STA

2
1 – 20

Read Stability – Defines the number of sequential samples within the motion band for
which a reading is regarded as stable; defined in conjunction with MOTBAND and
RD-MOT parameters. When the applied weight has settled within the stability range, a
rolling averaging filter is applied to improve the stability.

Table 3-7. Configuration Menu Parameters

GRADS

010000

CONFIG

FORMAT

CALIBR

PROGRM

DIG IN

ALGOUT

PASWRD

TEST

MENU

0 dd

ZTRKBN

ZRANGE

2 %

INIZR

0 %

MOTBAN

1 dd

RD-STA

002

RD-MOT

001

OVRLOA

FS-2

FS-1D

FS-9D

FS

DSPRAT

0.10SEC

0.25SEC

0.50SEC

0.75SEC

1.00SEC

NONE

SMPRAT

5 HZ

10 HZ

20 HZ

40 HZ

DIGFIL

ON

OFF

DFSENS

LIGHT

MEDIUM

HEAVY

DFTHRH

000000

TAREFN

BOTH

NOTARE

PBTARE

KEYED

Note