4.2 Parts

Counting

Use this application to count samples of uniform weight. Three different modes can be selected:
Standard Counting, Check Counting, or Fill Counting.

4.2.1 Standard Parts Counting (default)

In the upper left portion of the home screen, select Counting.
The default (or last) Average Piece Weight (APW) is displayed.

Place objects on the pan to display the number of pieces.

The PARTS COUNTING-
Standard Home screen

Main Display Line
Second Display Line

Reference Fields

Application Buttons

Functions

Application Icon

The Auto Optimization feature improves counting accuracy:
it automatically re-calculates the piece weight as parts are
added. By default APW Optimization is Off.
When a new APW is established, APW Optimization occurs
when the number of pieces added to the pan is between one
and three times the number already on the pan. The screen
momentarily indicates Auto Optimization in the information
line.

4.2.1.1 Edit Settings: To view or adjust the current settings, touch the Edit Settings button:

The Settings screen is now displayed.

Settings available: APW, Sample Size,
Functions available: Recalculate APW,
Return to Application

Establish an Average Piece Weight (APW)
Each time a new type of part is counted, the nominal weight of
one piece (Average Piece Weight or APW) must be established
using a small quantity of pieces. This APW is stored until
replaced by another APW.

There are two methods to establish the APW value:

1. The actual APW is known; or

2. The APW must be derived by weight. For this case the

current sample size will be used

To choose a different value, establish it beforehand, using the Sample Size button. (See below.)