EXPLORER BALANCES

EN-67

4.10.6 Measuring the Density of Porous Material Using Oil

To enable this feature, enter the Density Determination
Setup menu, and set the following:

9

Density Type: Solid

9

Liquid Type: Water

9

Porous Material: On

The DENSITY DETERMINATION – POROUS Home screen

Main Display Line
Second Display Line

Reference Fields

Application Buttons

Functions Application Icon

Confirm the default values displayed (Water Temp) are correct.
To edit the default values, touch Edit settings.
The Settings screen appears.

Settings available: Water Temp, Oil Density
Functions Available: Return to Application

The balance calculates water density based on the water
temperature value entered (look-up table).
Measure the actual water temperature using a precision
thermometer.
To adjust the Water Temperature or Oil Density values,
touch the Water Temp or Oil Density button.

Numeric input windows appear.

Key in the desired value,
then press 9.
The display returns to the previous
screen with the new value highlighted.

To return to the Density Determination
home screen, touch
Return to Application.