Rice Lake Explorer Series Toploader, Ohaus User Manual

Page 80

Advertising
background image

EN-78

EXPLORER

BALANCES

To adjust the water temperature, touch the Water Temp button.

The balance calculates water density based on the water
temperature value entered. Measure the actual water
temperature using a precision thermometer.

A numeric input window is displayed.

Key in the desired temperature, then press Done.

The display returns to the previous screen with the new value
highlighted.

Continue to enter the required information
following the above procedure:

Pipette Inaccuracy, via the Inaccuracy button

Barometric Pressure, via the Barometric Pressure button

Pipette Imprecision, via the Imprecision button

Pipette Name, via the Pipette Name button

Pipette Number, via the Pipette number button

When finished, press Return to Application.
The Application home screen appears, and you can begin the
Pipette Adjustment process.